----- Supply network analyzer CVM-144 ------ User's manual --- Page No. 8
4.2.- CVM-144 connection terminal arrangement (power supply connection terminal)
(see lable on the rear part)
9
12
56
7
83
4101112
S2 S1 S2 S1 S2 S1
L1
P2
P1
L2
P2 P1
L3
P2 P1
Power Supply
230 V
N
L1
V
L2
VV
L3
CAT III
300V
P
-N
-P
520V
I
n
/
5A
300 V
No. Terminal description No. Terminal description
1
2
3
4
5
6
*Power supply AL1
*Power supply AL2
IL1 S2 Current measurement
IL1 S1 Current measurement
IL2 S2 Current measurement
IL2 S1 Current measurement
7
8
9
10
11
12
IL3 S2 Current measurement
IL3 S1 Current measurement
Neutral
VL1 Measurement
VL2 Measurement
VL3 Measurement
NOTE: Terminals 3, 5, 7 are internally connected to the 9 terminal (Neutral)
... / 5 A current inputs are isolated for the ITF model.
* Lower connection terminal (SDC Model)
Descripción Borne
1
2
Supply voltage (+) d.c.
Supply voltage (-) d.c.