----- Supply network analyzer CVM-144 ------ User's manual --- Page No. 32
a.- If an ENERGY parameter is chosen: kWh (31), kvarhL (32) or kvarhC(33)
OUT 1 RELAY 1
PULS
xxxx
kW / pulse (1)
(1) Value of energy in kW : four digits with floating decimal point
Set-up procedure:
- "max" key: to modify the value of the blinking value. Every time it is pressed the
value is increased.
- "min" key: to validate the blinking value and go to the next digit.
NOTE : When the last digit is reached, the position of the decimal can be move point
with the "max" key.
Example for setting a 500 W / 1 pulse:
Firstly we enter the value, 0500, and following we place the decimal point at
the right position with the "max" key 0.500 kW.